Model: ABL-8L
Type: All -in one, microprocessor controlled
Test Standards: ISO 13736/ IP170, ISO 1516, ISO 1524
Measuring Range: -30°C to +80°C at cooling medium of 10°C, 0°C to +110°C at cooling medium of 40°C
Measurement Modes: ISO13736 Normal, ISO13736 SPE (search), ISO13736 Rapid, ISO1516, ISO1523, and User Custom
Bath: Metal Block
Sample Cooling/Heating: By Peltier Device
Display: 5.7 inch (117 x 88mm) color LCD with LED back light
Stirring: 0, 30 rpm (selectable)
Temperature Sensor: PT-100 in stainless steel sheath
Flash Detector: CRC Thermocouple
Ignition Source: Gas ignition with automatic lighting or electric ignition. Interchangeable
Barometric Correction: Automatic correction by built-in barometric pressure sensor
I/O Port: 1x RS-232C (for PC or Optional Printer); 1x USB channel (for USB keyboard or USB Flash Memory)
Safety Shutdown: Automatically shuts off and the problem is reported by buzzer and display, in case: (a) EFP+10 oC or 70 oC (ISO13736)110 oC(ISO1516,ISO1523) is reached, (b) temperature sensor is found defective, (c) flash detector is found defective, (d) thermofuse is blown, (e) electric Igniter is blown, (f) built-in battery is found drained out, (g) test cover is not set in place, (h)control computer runs away (no display)
Gas Supply: (When Gas Ignition is used), LP gas or natural gas (Max. pressure:10kPa)
Power Supply: 100/120VAC or 220/240 VAC, 50/60Hz
Power Consumption: 250VA MAX.
Electricity Consumption: 30Wh for 1 test (about 30 min), 11.34gCO2 (@ 0.378kgCO2/1kWh)
Ambient Temp. Range: 10°C to 35°C
Dimensions & Weight: 230mm (W) x 470mm (D) x 390mm (H) 16kg
Requirements For Chiller Option For Abl-8L
Type: Circulator with heating/cooling capacities. Open bath type.
Temperature Range: 5°C to +40°C or wider
Pump Capacity: Pressure:0.2bar minimum; Flow rate:2L/min minimum
Cooling Capacity: 250W@10°C minimum
Heater Capacity: 500W minimum
Bath Volume: 3L minimum
IQ/OQ/PQ Qualification: Yes, possible