Model: MPC-602
Type: Mini Pour Point(PP) and Cloud Point(CP) tester with 6 test heads in Bench-top package. With sequential CP and PP measurement capability. Sample cooling and pre-heating by Peltier modules with external cooling liquid.
Test Standards: ASTM D6749/D97, ISO 3016(PP), ASTM D7683/D2500, ISO 3015 (CP)
Specimen Volume: 4.5ml
Measuring Range (Typical*): -65°C to +51°C to with cooling liquid of -35°C, -60°C to +51°C with cooling liquid of -25°C; -40°C to +51°C with tap water of 20°C, *:Sample viscosity, etc. affects on lowest temperature of the measuring range.
Measurement Modes: Selectable from various modes. 1.CP mode (0.1 or 1.0°C, Selectable); 2.PP modes: Programmed by the user. Programmable parameters are: *Amount of applied air pressure for PP detection, to accommodate different sample types: L(low) for diesel fuels, H(high) for lube oils, VH(very high) and UH (ultra high) for residual fuels and similar samples. *Testing intervals: 1.0°C, 2.5°C, or 3.0°C (In total, 4x3=12 modes for PP.); 3.CP/PP modes: CP is determined and then PP. PP detection is programmable by the user with the same parameters as PP modes. (12 modes in total.)
Sample Automatic Pre-Heating: Automatic preheating at either +45°C or EPP+10°C. (EPP = Expected Pour Point)
EPP Setting: Test parameters, EPP, bath temperature, sample temperature, PP, and CP displayed on VFD. Temperatures displayed with 0.1°C increments.
Printer: (built-in, thermal type). PP, CP, mode, EPP, etc. printed by thermal printer.
Epp Setting: EPP(Expected Pour Point) needs to be set prior to test starts.
Specimen Cup: Cylindrical glass test jar with 4.5ml sample volume.
Sensors: Compound type sensor assembly for PP and CP. PP detected by air pressure method (patented). CP detected photo-electrically. PT100 temperature sensors.
Sample Cooling Rate: As standard, 4°C/min. till EPP+40°C, and 1°C/min. Thereafter, cooling process is programmable.
Safety Shutdown: As hot side of Peltier module reaches 60°C while preheating, warning buzzer beeps and heating stops.
Data Output: RS-232C 1 channel
Power Requirements: 100, 120, 220, or 240VAC 1.5kW
Size (W x D x H): 800W x 550D x 850H, 100kg
IQ/OQ/PQ Qualification: Yes, possible